Heracell 150i/240i 50115191_C_01-2018 112 Thermo Scientific
Chapter 7 Handling and control (iCanTM touchscreen controller)
7.14.3.3. Setting the gas-tight screen:
As the aperture cross-sections are smaller when accessing the samples, devices equipped with
the optional gas-tight screen achieve shorter recovery times for the incubation parameters:
Work space temperature,
•CO
2
concentration,
•O
2
concentration,
relative humidity.
After conversion of the device, the device control system must be adjusted to the gas-tight screen
option.
Figure. 54. Gas tight screen
1. To toggle between two options [1]:
Press the + key [4].
Note
Low humidity:
The enabling/disabling of the Low humidity function is entered into the event list.
If the Low humidity function is enabled, the corresponding icon [5] is displayed in the main menu.
Note
Malfunction:
The conversion to gas-tight screen causes a change of the control parameters.
Failure to set the gas-tight screen function in accordance with the actually installed door version may
impair the incubation performance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Heracell 150i and is the answer not in the manual?

Thermo Scientific Heracell 150i Specifications

General IconGeneral
BrandThermo Scientific
ModelHeracell 150i
CategoryAccessories
LanguageEnglish

Related product manuals