EasyManuals Logo

Thermo Scientific Heracell 150i User Manual

Thermo Scientific Heracell 150i
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Heracell 150i/240i 50115191_C_01-2018 95 Thermo Scientific
Chapter 7 Handling and control (iCanTM touchscreen controller)
The auto-start-instruction menu [1] is displayed.
2. To exit the auto-start-instruction menu and cancel auto-start:
Press the END key [4].
3. Activate the auto-start routine:
Press the START key [3].
4. To air the work space, open both device doors.
5. When the audible alarm sounds after 30 seconds, close both device doors.
The trend status indicator [5] is displayed.
6. During the progress, the status indicator displays the following information:
Temperature,
start time.
Note
Cancellation:
The auto-start routine can be canceled at any time!
Press the Stop key [6].
Automatic restart:
The auto-start routine is restarted automatically if the routine is canceled due to one of the following
events:
The glass door is opened,
the outer door is opened on a device with optional gas-tight screen,
the power supply is interrupted.