Chapter 7 Handling and control (iCanTM touchscreen controller)
or
Press the - key [2].
2. To accept and save the change:
Press the ENTER key [3].
The system returns to the Options menu.
7.14.3.4. Switching the water level sensor on/off:
For incubation operation with ambient humidity or if the auto-start routine is to be run dry (without
water), the water level sensor can be switched off.
This prevents alarm messages for the water level sensor by the device-integral monitoring system
from being issued.
Figure. 55. Water level setting
1. To toggle between two states [1]:
Press the + key [4].
or
Press the - key [2].
2. To accept and save the change:
Press the ENTER key [3].
The system returns to the Options menu.
7.14.3.5. Switching the audible alarm on/off:
If the device-integral monitoring system detects an error:
• an audible alarm sounds
• in addition to the visual error message and the switching of the alarm relay.