Heracell 150i/240i 50115191_C_01-2018 155 Thermo Scientific
Chapter 10 Maintenance
Measurement example:
•CO
2
set value: 5 %
Reference temperature: 5.6 %
1. Press the CO
2
DISPLAY key [1].
The CO
2
menu is displayed.
2. To exit the CO
2
menu:
Press the END key [3].
3. To enter the Calibration submenu:
Press the CALIBRATION key [2].
4. To enter the measured value (targeting value):
The targeting value can be increased or reduced in increments; if you keep the + key [4] or the
– key [6] depressed, the function switches to a rapid increase/reduction; after approx. 3
seconds, another increase/reduction occurs.
To increase the targeting value:
Press the + key [6].
To reduce the set value:
Press the key [4].
5. To accept and store the targeting value:
Press the ENTER key [5].
The system returns to the main menu. The temperature display shows the current actual
value measured in the CO
2
display.
10.9. Replacing the gas inlet filter
The gas inlet filter (CO
2
-/O
2
-/N
2
supply) has plastic threads and is screwed by hand into the
threaded hole at the control box.
Note
Excessive CO
2
content:
Excessive CO
2
content after the calibration can be reduced by leaving the device doors open for approx. 30
seconds.
Value reset:
If the value isn’t changed within the next 30 seconds, the system automatically exits the menu, and the
most recently confirmed value is preserved.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Heracell 150i and is the answer not in the manual?

Thermo Scientific Heracell 150i Specifications

General IconGeneral
BrandThermo Scientific
ModelHeracell 150i
CategoryAccessories
LanguageEnglish

Related product manuals