EasyManuals Logo

Tomey AL-4000 User Manual

Tomey AL-4000
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #157 background imageLoading...
Page #157 background image
4-5
C2 : Anatomic anterior chamber depth; distance
from the corneal vertex to the iris surface
(mm)
= 0.56 + Rag
4/
22
AGRag
r < 7 Rag = 7
r 7 Rag = r
AG = 12.5L / 23.45
AG > 13.5 AG = 13.5
3. Personal SF
22/4{
2
CAQCQAQBQBQSF
Where
AQ = (nc-1) - 0.001AREF{V(nc - 1) - r}
BQ = 0.001AREF{L2V(nc - 1) - r(L2 - Vna)}
-{L2(nc - 1) + nar}
CQ1 = 0.001AREF[V{nar - L2(nc - 1)} + L2r]
CQ2 = 1000na{nar L2(nc - 1) - CQ1}/P
CQ3 = L2nar-0.001AREFL2Vrna
CQ = CQ3-CQ2
AREF : Refractive power of eye after surgery (D)
4. Corresponding SF
SF = 0.5663A-65.60
Where
A :A-Constant