5.4.2 Cleaning and Disinfection of measurement probes (for Europe)
[Specific Section for European continent]
Follow the legal requirements and regulation
in the country.
Refer to the instruction manual concerning
test-fluid for detail.
a) Cleaning
Holding and pulling the cord may damage the
inner core wires.
Do not clean the connecting section of the
connector.
When cleaning the main body of the measurement probe and, wipe
it with a suitable fluid eg.Cleanisept® Wipes
and then wipe with
cotton dampened with water
Wipe the cord with a well wrung cloth dampened with water and
remove water after cleaning.
b) Disinfection
Effective chemicals vary depending on the types
of germs and viruses. Select appropriate
chemicals and immersion time based on the
guidelines, articles, literature, etc. provided by
local administrative organizations and/or
societies.
A high-pressure steam sterilizer is acceptable
only for the immersion attachment.