5-5
When sterilizing the pachymetry probe with EOG, remove the
protective cap from the tip of the probe and put the whole probe in a
sterilizing pack.
<Sterilization conditions>
EOG: Mixture of ethylene oxide gas (20%) and carbonic acid gas
(80%)
Sterilizing temperature: 50 ± 5°C
Sterilizing pressure: (9.8 +/- 0.98) x 104 Pa
Sterilizing time: 5 +/-1 hours
Aeration time:
Corneal thickness probe: 10 hours or more (continuous)
Immersion attachment: 10 hours or more (continuous)
Disposable tip: 20 cycles of 30 minutes (supply
air for ventilation after each
cycle)
* Use the probe after leaving for at least 7 hours, and the disposable tip for at
least 6 days, after completing sterilizing with EOG.