i. Important safety information
ڦ
Do not install this instrument in a location where explosives or inflammable
substances are used or stored. Fires or explosion may occur.
ڦ
Do not remove the cover of the instrument. You may be directly exposed to
high voltage sections.
ڦ
Do not disassemble or modify the instrument. You may be directly exposed to
ڦ
Disconnect the AC cord from all the connected devices before installing and/or
servicing the instrument. Otherwise, you may get an electric shock.
ڦ
Sterilize measurement probes before use.
ڦ
Do not use a measurement probe with a damaged contact section.
Measurement error may occur, as well as the corneal or eyelid may be injured.
ڦ
Immersion attachment and disposable tip are a disposable part. Do not reuse
these parts. Otherwise, you may contract diseases.
ڦ
You are cautioned that changes or modifications not expressly approved by
the party responsible for compliance could void your authority to operate the
ڦ
Do not place water or chemicals on the instrument. Any water or chemicals
entering the instrument may cause an electric shock or failure.
ڦ
Do not allow water or chemicals to splash on measurement probes except for
the tips. Any water or chemicals entering the probes may cause an electric
shock or failure.
ڦ
Only use the specified terminal for connection of the instrument. Using
another type of terminal may result in failure of the instrument.
ڦ
This instrument is a measuring device specially designed for ophthalmology.
Never use the instrument for other purposes.
ڦ
The terminal for connecting the instrument to external devices is not isolated
from the internal circuit. Inappropriate wiring may damage the internal circuit.
Never touch any of these terminals and the patient at the same time. Be sure to
contact Tomey or our local distributor before using the external output
terminal.
ڦ
When operating this instrument connected to other devices, only use devices
that satisfy IEC60601-1 or equivalent safety requirements, or that conform to
IEC60950 and whose power source is insulated with an insulation transformer,
in order to satisfy the relevant safety requirements for medical equipment.