47
either or button to change the settings
so that the secondary display shows “Set”.
5. Press the UTIL button to check whether other
desirable setups have already been set. With
all settings done, exit the menu by long press
of the UTIL button.
6. The meter has now saved all current settings
into internal memory. At next power-up, the
meter will turn on and recall the saved
settings.
NOTE: The meter allows one set of settings to be
stored into memory. Therefore, the same
procedure is used to overwrite previously stored
settings into memory.
Prevent Overwrite of Stored Settings
In the utility menu, the PuPoption default setting is
always PrE. If it is required to overwrite previously
stored settings for power-up state, the option should
change to Set”. Therefore, when entering the utility
menu, be sure not to change to Set” to prevent
overwriting any previously stored power-up settings.