50
In “dEF” option, push either or button to
choose “NO” or “yES”. “NO” means the instrument will
not be set back to default. “yES” indicates to reset all
settings to default and to clear previously stored setup.
Default Setting: No
NOTE:
Exiting mode also decides whether or not to
perform “yES” function. At the use of “exiting without
saving”, “yES” is ineffective; when “saving and exiting”
is selected, restore operation will be efficient.
NOTE: In the case where “PuP” is set to “Set” and
“dEF” is set to “yES” simutaneously, the “PuP” setting
has priority over the “dEF” setting. This means the
instrument will not be set back to default upon saving
and exiting the utility menu. Instead, the power-up
settings will be stored and will be recalled upon the
next power-up of the instrument.
Indicate battery voltage(bAtt)
When menu option changes to “bAtt”, the secondary
display will indicate battery voltage that is for
reference instead of for operational function.