ALADDIN - User manual Rev. 19 EN of 10/07/2019
14 CHANGING THE FUSES
Step 1
Open the fuse box cover
using a screwdriver
Step 2
Take out the fuse box (use
a screwdriver to release it)
Step 3
Remove the blown fuse
from its seat and replace it
with an identical one, as
indicated in the table
below and on the
instrument label.
Step 4
Push the fuse box carefully
back into position
It is mandatory to use fuses only with the indicated characteristics
The use of undersized fuses can cause the interruption of power to the device during normal working
conditions. In this case there is no risk to the user, nor for the patient, but the device turns off at inopportune
moments, and this can cause data loss
The use of oversized fuses can lead to damage to the internal electronics of the device due to current
overload for non-interruption by fuses. In this case you do not identify risks to the patient, but on the user
or damage to the device and then stop working and possible data loss.
T 2.5 A L 250 V anti-surge