ALADDIN - User manual Rev. 19 EN of 10/07/2019
8
PUPILLOMETRY - It is performed with image analysis of the sequence of acquisition. The acquisition is
performed with LEDs of different wavelengths. The device uses infrared LEDs to dilate the pupil and white
LEDs to reproduce photopic light conditions and to contract the pupil. There are three modes:
DYNAMIC Pupillometry acquisition in dynamic controlled light condition: mesopic photopic
mesopic.
PHOTOPIC Pupillometry acquisition in static photopic controlled light conditions.
MESOPIC Pupillometry acquisition in static mesopic controlled light conditions.
IOL CALCULATION - Intraocular lens power calculation and suggestion made using the following formulae:
Holladay 1, Haigis, Hoffer Q, SRK / T, SRK II, Camellin-Calossi, Shammas No history, Barrett Universal II. There
are three modes:
IOL calculation: Calculating the total spherical power for an intraocular lens to be implanted.
Toric IOL calculation: Calculating the spherical equivalent power, cylinder power and placement axis
for a toric intraocular lens to be implanted.
Post refractive IOL calculation: Calculating intraocular lenses for patients who have undergone
refractive surgery to correct myopia or hypermetropia.
The ALADDIN HW3.0 has an onboard PC with a dedicated software providing all the functionalities described
above.
Please refer to the literature references listed in 19. Appendix: bibliography.
1.5 Essential Performance
Keratometry (KER) Measurement
Axial Length (AL) Measurement
Anterior Chamber Depth (ACD), Central Corneal Thickness (CCT), Lens Thickness (LT) Measurement
1.6 Interaction with the patient
Patients do not control the device. They are positioned with their chin on the chinrest and their forehead
on the forehead-rest and they are asked to stay perfectly still and look at the fixation point with one eye.
The device is entirely controlled by specialized personnel.
The parts in contact with the patients are the followings:
ABS chinrest
Teflon forehead rest

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon ALADDIN HW3.0 and is the answer not in the manual?

Topcon ALADDIN HW3.0 Specifications

General IconGeneral
BrandTopcon
ModelALADDIN HW3.0
CategoryMedical Equipment
LanguageEnglish

Related product manuals