Before turning on the device, remove the cover. Never put the cover on when the device is on.
4.7 Checking the measurements
The calibration must be checked when the device has been transported from one place to another
and when it has suffered an impact or thermal shocks.
Check the measurements every day when turning on the device using the instrument provided.
The user of the device must check that the measurements provided by the device are plausible.
It is advisable to visually check all the light sources before examining patients to make sure that they
come on properly.
If the device frequently emits error signals, turn it off and contact technical support to have the device
checked.
In patients with blue eyes, acquisition of pupillometry in mesopic lighting conditions can be difficult
to accomplish. In this case, we suggest acquiring the mesopic data through dynamic pupillometry.
Contact lenses must not be worn by the patient during data acquisition.