EasyManua.ls Logo

Topcon CA-200F - ACQUISITION ENVIRONMENT

Topcon CA-200F
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CA-200F Corneal Analyser - Rev. 7 16/01/2012
22
ACQUISITION ENVIRONMENT
The acquisition environment is divided into three sections: Topography, Pupillometry and Fluorescein.
Each time you access acquisition, the instrument is set to topography (fig. 1-8).
fig. 1-8
The following buttons are available in the acquisition window:
R and L: indicates the eye being acquired, highlighted in yellow.
TOPO, PUPI, FLUO: indicates the type of acquisition (highlighted in yellow); you can select from
topography, pupillometry and fluorescein.
Acquisition Gallery
A preview of the acquired image is shown in the acquisition gallery (fig. 1-9, fig. 1-10, fig. 1-11 for
topography, pupillometry and fluorescein, respectively).
fig. 1-9
fig. 1-10
fig. 1-11

Related product manuals