35
BEFORE REQUESTING SERVICE
BEFORE REQUESTING SERVICE
CAUTION MESSAGES
CHECK ITEMS
CLEAN THE COVER GLASS Carry out cleaning of the cover glass.
Refer to the instruction at page 34.
DIOPTER OVER
PRISM OVER'
ERROR
Check to see that the lens is in the measurable scope.
Check to see that the lens is free from any flaw, dust or oil.
Clean both glasses and turn power on again.
Check the lens value is not beyond the unit's measuring
range.
INITIAL ERROR Request repair service.
Remove the lens from the lens support, and turn on power
again.
PAPER END
(with printer specification)
Printer paper is out. Load new paper.
CLOSE PRINTER COVER
(with printer specification)
Close printer cover. Refer to the instruction at page 16.
The instrument does not get ready for
operation even if the power switch is
turned on.
Re-plug the power cord.
Give a check to the fuse.
The monitor screen can not be clearly
seen.
Adjust the BRIGHT control dial.
S, C values are wrong. Is the lens place with power on?
Remove the lens and turn on power again.
Is the measuring beam blocked by, dust, marks,
grease, etc. in the measured lens?
Marking is poor. Replace the marking ink cartridge. For a lens with
sharp surface curve, use the optional steel needle
marking set.
The screen went out all of a sudden. The auto shut-off function is on.
Press the MEMORY button , and the instrument will
resume.
Pushed the Print button but the printer
does not work. (with-printer type)
Is printer paper set properly?
Not inside out?