30
8
Turn Off the instrument by pressing the Power key.
9
Turn the power ON again and follow the checking procedure again to conrm that proper
calibration has been completed.
10
Repeat the checking and, if necessary, the adjusting procedure for the Y-axis.
11
If the Y-axis requires adjustment, begin at Step 1 on page 26. However, when turning on
the instrument in calibration mode, press and hold the Y AXIS button while pressing the
Power key. This activates the Y-axis calibration mode.
Press
Press and
Hold
For Y-axis
calibration