Congratulations !
You have purchased the latest in Handheld pH-mV-Temperature
instrumentation. We trust that your new WP-80 will give you many years
of reliable service.
The WP-80 is a breeze to operate. This manual has been designed to
help you get started, and also contains some handy application tips. If at
any stage you require assistance, please contact either your local TPS
representative or the TPS factory in Brisbane.
The manual is divided into the following sections:
1. Table of Contents
Each major section of the handbook is clearly listed. Sub-sections
have also been included to enable you to find the information you
need at a glance.
2. Introduction
The introduction has a diagram and explanation of the display and
controls of the WP-80. It also contains a full listing of all of the items
that you should have received with your WP-80. Please take the time
to read this section, as it explains some of items that are mentioned
in subsequent sections.
3. Main Section
The main section of the handbook provides complete details of the
WP-80, including operating modes, calibration, troubleshooting,
specifications, and warranty terms.
4. Appendices
Appendices containing background information and application notes
are provided at the back of this manual.
Model WP-80
pH-mV-Temp. Meter
Version : 3.51
Date : 02/07/2003
Author : MS

Questions and Answers

  • H
    Hannah BennettSep 1, 2025
    How to fix temperature if it is inaccurate and cannot be calibrated on TPS WP-80 Measuring Instruments?
    • M
      Michael RobinsonSep 1, 2025
      If the temperature reading is inaccurate and cannot be calibrated, there may be a faulty connector or temperature sensor. Check the connector and replace if necessary. If the problem persists, fit a new temperature sensor, part number 121247.
  • F
    Frederick RichardAug 30, 2025
    What to do if my TPS WP-80 unit fails to calibrate, even with a new sensor?
    • C
      Catherine TaylorAug 30, 2025
      If the unit fails to calibrate even with a new sensor, the calibration settings may be outside allowable limits due to a previous failed calibration. Initialise the unit.
  • T
    toddcourtneyAug 29, 2025
    Why does my TPS WP-80 display “Man” when the temperature sensor is plugged in?
    • J
      Jeremiah GillAug 29, 2025
      If the display shows 'Man' when the temperature sensor is plugged in, it could indicate a faulty temperature sensor. Try fitting a new temperature sensor, part number 121247.
  • S
    sharon99Aug 28, 2025
    What does it mean if my TPS WP-80 Measuring Instruments display flashes “ATC” and “LIMIT”?
    • S
      Steven JohnsonAug 28, 2025
      If the display flashes “ATC” and “LIMIT”, the temperature is not within the ATC limits. Cool or heat the solution before taking measurements.
  • B
    Benjamin MaldonadoAug 27, 2025
    Why are my TPS WP-80 readings inaccurate, even after calibration?
    • J
      Juan GoodwinAug 27, 2025
      Inaccurate readings after calibration may be due to a blocked reference junction. Clean the reference junction as per the instructions supplied with the sensor.
  • K
    Kyle CamachoAug 27, 2025
    What to do if my TPS WP-80 meter displays the word OFF, and switches off?
    • K
      kennethwilliamsAug 27, 2025
      If your meter displays 'OFF' and switches off, the battery is likely below 5.00 volts. Recharge the battery. If this doesn't work, check the charger and replace the battery if the charger is OK.
  • J
    Jenna PearsonAug 25, 2025
    What to do if my TPS WP-80 battery does not charge?
    • K
      Kim VasquezAug 26, 2025
      If the battery isn't charging, it could be a faulty charger or battery. Connect the charger and switch the power on. Display the battery volts in the battery saver menu. If the battery volts are increasing, the charger is OK. If the battery volts do not increase, then the charger is faulty. Replace the charger or the battery, as required.

Summary

Introduction

WP-80 Display and Controls

Details the display and physical controls of the WP-80 instrument.

Specifications

Provides detailed technical specifications for the WP-80, including ranges, resolution, and accuracy.

WP-80 Menu Structure

Operating Modes

Selecting pH, mV or Relative mV Mode

Explains how to select between pH, mV, and Relative mV operating modes on the WP-80.

pH Calibration

Calibration Procedure

Step-by-step guide for performing a pH calibration on the WP-80 instrument.

Calibration Notes

Provides important notes and considerations for performing accurate pH calibrations on the WP-80.

Relative mV Calibration

Temperature Calibration

Calibration Procedure

Step-by-step guide for calibrating the temperature measurement function of the WP-80.

Good Laboratory Practices (GLP)

Failed Calibration

Describes the GLP function's behavior when a calibration fails and how it affects stored data.

Notepad Function

Automatic Datalogging

RS232 Port

Selecting Buffers for Auto Buffer Recognition

Initialising the WP-80

Troubleshooting

General Errors

Lists common error messages, their possible causes, and remedies for the WP-80.

pH and mV Troubleshooting

Addresses specific troubleshooting steps for pH and mV measurement issues with the WP-80.

Temperature Troubleshooting

Provides troubleshooting guidance for temperature-related inaccuracies or calibration failures on the WP-80.

Appendices

Determining if an instrument or electrode is faulty

Provides a diagnostic test to help determine if the WP-80 instrument or its electrode is faulty.