23D01-060en202008 Manual nanoFlu
Use
Malfunction &
Maintenance
Commis-
sioning
Introduction
General
Information
FAQ Technical DataWarranty
Customer
Service
Contact
Keyword
Index
Accessories Calibration
nanoFlu // Calibration
All TriOS controllers have the ability to set scaling factors and oset values for the measurement parameters.
Please refer to the appropriate manual. Make sure not to carry out double scaling with the sensor: once in the
G2 sensor menu directly and once more with the TriOS controller!
Customer calibration can be used as a ne adjustment of the sensor for special media and is not intended to
replace the manufacturer calibration.
1
2
Laboratory value for the test substance [µg/L]
Measured concentration [µg/L]
Measurement ranges and detection limits of the scaled parameters are dependent
on the scaling factor!
NOTICE