EasyManua.ls Logo

triOS nanoFlu - Page 24

triOS nanoFlu
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 D01-060en202008 Manual nanoFlu
Use
Malfunction &
Maintenance
Commis-
sioning
Introduction
General
Information
FAQTechnical Data Warranty
Customer
Service
Contact
Keyword
Index
AccessoriesCalibration
Calibration // nanoFlu
4. The scaling factor can be calculated using the following equation
scaling factor =
With lab for the laboratory values and measuredvalue for the values output by the sensor.
For the previous example in the gure, this means:
scaling factor = = 3
5. If a non-contaminated sample is not available, at least two samples with very di󰀨erent levels of contami-
nation are needed. In this case, you will rst calculate the scaling factor.
scaling factor =
Calculation of the o󰀨set without zero-point measurement (1.):
oset = measuredvalue2 -
measuredvalue2 should be signicantly larger than measuredvalue1. The o󰀨set can also be found using
the abscissa of the straight line (intersection with X axis). For the above example, this means:
scaling factor = = 3
oset = 40 - = 40 - 30 = 10
(lab2 - lab1)
(measuredvalue2 - measuredvalue1)
lab
measuredvalue2 - oset
90 mg/L
(40 - 10) mg/L
lab2
Scaling factor
90
3
90 - 30
40 - 20
Offset = 10 (Abszisse)
1
2
Laboratory value for the test substance [µg/L]
Measured concentration [µg/L]
O󰀨set = 10 (abscissa)

Related product manuals