41D01-060en202008 Manual nanoFlu
nanoFlu // Annex
Write single register (0x06)
With the “write single register” function, instead of changing the conguration values, a single measurement
trigger is sent, in which a value ≠0 is written.
Designation Address Description
Trigger measurement 1 A single measurement is taken.
Report slave ID (0x11)
Provides the sensor designation followed by the serial number followed by the rmware version each as a
null-terminated ASCII character string.
Example:
n a n o F l u 0x00 D 2 0 0 0x00 1 . 0 . 0 0x00
Writing to the conguration registers should be done as seldom as possible and
especially not in every measuring cycle, otherwise the ash memory may be
damaged.
NOTICE
Parametername RW 603 Char[32] Name of the parameter
Temperature R 998 Float
Sensor temperature in [°C] during the last
measurement.
Concentration / scaled
concentration
R 1000/1500 Float