86 D01-051en202111 Manual TriBox3
Annex // TriBox3
Spectral data
These are pairs of wavelengths and Y-values: A wavelength value is always accompanied by the corresponding
Y-value.
6.3 Registers of the TriBox
Register range Datatype Read/Write Description
20 … 39 Char[40] R Name of Tribox: Tribox_9401
100 … 101 int32 R Serial number
102 … 107 DateTime RW System time
108 Char[64] RW User-dened name of the TriBox
6.4 Register of other multiparameter probes
Where possible, all devices have the same register mapping. The following table lists the registers used by all
sensors with up to 32 measurement parameters.
Register range Datatype Read/Write Description
20 … 39 Char[40] R Name of the device
80 Int16 R Number of parameters
81 DateTime R Time of last measurement
1000 R Data of up to 32 measurement parameters
1000 … 1063 32 Floats R Up to 32 measurement results
1064 … 1383 32 * Char[20] R Names of the parameters in the order of the meas-
urement results
1384 … 1639 32 * Char[16] R Names of the units of the parameters in the order of
the measurement results
At the moment the TriBox sets the Hi and Lo of the oats to the low endian format. This dif-
fers from the behaviour of the sensors. This may be changed in future software versions.