42 D1
+
PERSONAL DUST MONITOR
XD1+_User manual_P5650.1600_Rev A
The routine must be carried out using the Trolex Compliance Pack
+
required to conduct the
procedure, including:
Reference particle sizes (Reference material)
Dosing bottle
Sized reference material is used to periodically test specific operational aspects of the
D1
+
during the compliance audit check.
Once the D1
+
is connected to the BreatheLITE software, and the compliance audit test
has been selected from the maintenance menu follow the on-screen instructions to dose the
device with reference particulate material.
Note: The dosing bottle is a dispersing mechanism for the sample particulate material. Please
ensure that these components are kept in a clean and dry environment, free from moisture and
contaminants.
It is important to ‘prime’ the dosing bottle by shaking it vigorously to ensure the reference
material is free to disperse into the hood.
To dose the reference material into the particulate dispersion hood, use the built-in pumping
mechanism several times to active dosing.