Appendix A–12
With the source rod in the SAFE (shielded) position, place the
gauge on its side.
Clean the heads of the four corner screws that hold the bottom
plate to the gauge base (Figure A–3). Using a Phillips
screwdriver, remove the four screws in the corner of the plate
and remove the plate.
To reduce radiological exposure, stand to one side of the gauge.
Paying close attention to the position of the sliding block,
remove the block.
Using a stiff brush or rag soaked in alcohol, clean the sliding
block and the cavity.
Re-install the sliding block with the angled side up. Apply a
light coating of Magnalube-G paste to the top angled surface of
the sliding block.
Re-install the bottom plate. Do not over-tighten screws!
Ensure that the source rod moves up and down freely.
Figure A3. Cleaning the Tungsten Sliding Block
GAUGE BASE
CAVITY
SLIDING BLOCK
BOTTOM PLATE
FLAT HD
SCREWS (4)