EasyManuals Logo

Troxler EGauge 4590 User Manual

Troxler EGauge 4590
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
D-4
Sample analysis must be performed by a
licensed laboratory only.
WARNING!
Ensure that the source rod is in the SAFE position.
L
L
e
e
a
a
k
k
T
T
e
e
s
s
t
t
i
i
n
n
g
g
Routine leak testing of the Model 4590 is not required in the
USA due to the low activity of the source. However, if you
desire to perform a leak test of the Model 4590, please perform
the following using the Troxler Model 3880 Leak Test Kit (P/N
102868) and accompanying instructions.
1
1
.
.
Write the date, gauge model number, and serial number on
the sample form and label.
2
2
.
.
Ensure that the source rod is in the SAFE position.
3
3
.
.
Turn the gauge on its side and locate the opening where the
source rod extends through the gauge base.
4
4
.
.
Holding the wipe disk with tongs, wipe the area around and
inside the opening where the source rod extends from the
gauge base.
5
5
.
.
Pack the disk, as instructed, in the envelope and mail to
Troxler Electronic Laboratories, Inc. for analysis.
6
6
.
.
Secure the gauge properly.
NOTE

Table of Contents

Other manuals for Troxler EGauge 4590

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler EGauge 4590 and is the answer not in the manual?

Troxler EGauge 4590 Specifications

General IconGeneral
BrandTroxler
ModelEGauge 4590
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals