5-34 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
The Logging Options screen allows you to set up data logging options.
Figure 5-19 below shows the Logging Options screen.
Figure 5-19
Electrostatic Classifier Logging Options Screen
To log data, follow these instructions:
1. On the Setup screen, touch Logging Options.
2. Touch Enable Logging to allow data to be logged (default setting).
Inverted data is logged to a .txt file. If Enable Logging is unchecked,
sample data will not be saved in any format.
3. To specify a Delimiter for the inverted logged data, touch Tab,
Comma, or Semicolon. Tab delimiter is recommended to avoid
unintentional delimiting of any text in the Notes field which might use
commas or semicolons.
4. To specify a Decimal Symbol for the inverted logged data, touch Dot
or Comma.
Note: You cannot select comma for both the delimiter and the decimal
symbol. If you have specified a comma delimiter and then select
a comma decimal symbol, the delimiter changes.
5. Touch Log Raw Data to enable the logging of raw data in addition to
the inverted data. A .TIM file containing raw counts will be saved in
addition to the .txt file (see next section for details about .TIM files).
6. Touch Log to Internal Memory to log the data to the internal memory,
or leave unchecked to log to external USB drive (preferred). An on-
screen message tells you how much logging memory is available on
the destination drive. If box is unchecked and an external drive is not
detected, the message will report a value of 0.0.
7. Touch OK to save your settings or Cancel to cancel them.
Note: If Log to Internal Memory box is unchecked but no flash drive
is present when the start button is pressed, the classifier will
automatically check the box and switch to log to internal
memory.