A-4 Model 3082 Electrostatic Classifier and SMPS Spectrometer Model 3938
Environmental Conditions
Indoor use
Over-voltage category II
Pollution degree II
Ambient operating conditions ...
Temperature: 10°C to 40°C
Humidity: 0 to 90% RH, noncondensing
Storage conditions .................
Temperature: -10°C to 55°C
Humidity: 0 to 90% RH, noncondensing
Table A-4
Specifications of the Electrostatic Classifier with Long DMA
Particle type
Solids and nonvolatile liquids
Particle size range (Classifier mode)
Adjustable from 10 to 1000 nm
Sizing Accuracy
±1 % at 100 nm for 10:1 sheath/aerosol flow ratio
Maximum input concentration
10
7
particles/cm
3
Physical measurements - Long DMA Model 3081A
Height .................................
61 cm (24 in)
Outside diameter ................
7.6 cm (3 in) excluding ports
Weight ................................
5.5 kg (12 lb)
Monodisperse and
polydisperse aerosol ports .
¼-in OD
Sheath flow ports ...............
3
8-in OD
Complete instrument
Dimensions (LWH) .............
40.5 cm 28 cm 64.3 cm
(16 in 11 in 25.3 in)
Weight ................................
20 kg (45 lb)

Table of Contents

Other manuals for TSI Instruments 3082

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 3082 and is the answer not in the manual?

TSI Instruments 3082 Specifications

General IconGeneral
BrandTSI Instruments
Model3082
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals