Maintenance, Service, and Troubleshooting 6-25
Performing User Recalibrations and Checks
Check or recalibrate the sheath flow as part of your routing maintenance to
ensure ±2 percent accuracy of sheath flow rate. Full details of the sheath
flow calibration are given in Chapter 5, Performing a User Calibration.
The differential pressure sensor has some sensitivity to temperature
therefore the sensor be zeroed from time-to-time as well as before you
perform an impactor calibration.
Impactor loading can influence the accuracy of the impactor flow
measurement; therefore, check or recalibrate the impactor flow as part of
your routine maintenance. Full details of the sheath flow calibration are
given in Chapter 5, Performing a User Calibration.
Upgrading Firmware
Check the TSI web site periodically for upgrades to the Electrostatic
Classifier firmware and CPC firmware. Firmware upgrades are contained in
zip files available for download.
To find a specific firmware upgrade file, follow these instructions:
1. Open the TSI web site at http://www.tsi.com/.
2. From the Support tab > TSI Software and Firmware > Software and
Firmware Wizard.
3. Follow the onscreen instructions to find the correct firmware upgrade.

Table of Contents

Other manuals for TSI Instruments 3082

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 3082 and is the answer not in the manual?

TSI Instruments 3082 Specifications

General IconGeneral
BrandTSI Instruments
Model3082
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals