EasyManuals Logo
Home>TSI Instruments>Laboratory Equipment>3082

TSI Instruments 3082 User Manual

TSI Instruments 3082
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Unpacking and Setting Up 2-7
2. Tilt and lift off the cover leaving the five loosened screws in place.
3. Using a Phillips-head screwdriver, remove the two screws securing the
flag assembly and save for reuse.
4. Lift out the flag assembly by placing your fingers in the depressions
on either side and pulling up on the Polydisperse Flow outlet port.

Table of Contents

Other manuals for TSI Instruments 3082

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 3082 and is the answer not in the manual?

TSI Instruments 3082 Specifications

General IconGeneral
BrandTSI Instruments
Model3082
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals