2-12 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
6. Slide the neutralizer and flag assembly into the neutralizer housing in
the orientation shown in the photograph below. The Polydisperse Flow
outlet tube will fit into the cutout in the instrument case.
7. Reinstall the two retaining screws you removed in step 1.
The radioactive pop-up flag will be up, indicating that a neutralizer is
installed. The flow path inside the classifier has now been completed,
and aerosol pulled through the classifier inlet will now be exposed to
radiation.
C a u t i o n
The yellow indicator flag indicates a neutralizer is installed. A classifier
should never be shipped with a 3077/3077A neutralizer installed due to
transport regulations of devices containing radioactive material. See
Model 3077 manual for proper transport and shipping protocol.