Unpacking and Setting Up 2-21
3. Using a wrench, remove the Excess Flow (Sheath Flow ) fitting.
4. Remove any existing Teflon tape from the Excess Flow fitting. Apply
new Teflon tape to the fitting.
5. Replace the Excess Flow fitting on the Long DMA in a vertical
position.
Note: The fitting should be at a 90° turn from its original orientation.
6. Using a wrench, remove the Monodisperse Flow fitting.
7. Remove the protective cap from the new Monodisperse Flow fitting
and apply Teflon tape to the fitting.