C3 Submersible Fluorometer
998-2300 Rev. H Page 16
The temperature compensation coefficients for each fluorophore are
listed in the table below.
Fluorophore Coefficient
Rhodamine 0.026 / °C Exponential
Chlorophyll a 0.014 / °C Linear
Step 5: Save Calibration
After calibration is complete, select “Save Calibration”.
Data are now displayed as concentration estimates.
3.3.2 Raw Fluorescence Calibration – Blank Subtracted
Following steps 1-5 below will ensure values are reported as Blank
subtracted Relative Fluorescence Units (RFUB). See Appendix B for
recommended practices.
Step 1: Blanking
Place the C3 Submersible Fluorometer in a blank solution.
Note: A blank solution is a solution without the fluorophore of
interest (i.e. de-ionized water, artificial water, or filtered water).
Note: Ensure the optical head is free of air bubbles.
Wait until “Current RFU” readings have stabilized then select the “Set
Blank” button to set the blank value.
Wait for the blank value to be measured and set for all 3 gains.
Step 2: Selecting Units
Select RFUB from the units’ dropdown menu.
Note: Do not enter a value in the standard value box.
Step 3: Skip and go to Step 4
Step 4: Optional Temperature Compensation – Follow same steps
as the Direct Concentration Calibration procedure in Section 3.3.1.
Step 5: Save Calibration
After calibration is complete select “Save Calibration”.
Data are now displayed as blank corrected relative fluorescence
units.