C3 Submersible Fluorometer
998-2300 Rev. H Page 18
Atmospheric and Gravity Offset
Step 1 (Method 2):
Select the “Calibrate” button for depth in the “Current Data” screen.
Pressure in PSI will be displayed in the “Current Pressure” window.
After “Current Pressure” readings are stable, at desired location,
select “Set Zero”.
Step 2 (Method 2):
Select “Atmospheric and gravity offset”.
Enter the latitude.
Step 3 (Method 2):
Select “Save Calibration” button.
Depth values will now be scaled to the offset selected.
3.4 Log Setup Screen
The Log Setup screen allows the user to configure a logging session, set the
mechanical wiper to its home position, and enable logging. The C3
Submersible Fluorometer can log just over 64,000 lines of data.
Enter a log start date (MM/DD/YYYY) and time (HH:MM:SS).
Enter the sampling interval (HH:MM:SS). The minimum sampling interval is
1 second.
If the mechanical wiper is installed click the “Wiper” box to enable.
Select the number of revolutions from the dropdown box (1-10 revolutions).
All sensors will be wiped at specified revolutions before each measurement.
Note: If users select revolutions that exceed sampling interval a
warning message will be displayed.
Note: Do not click the Wiper Box to enable wiper if the wiper is not
installed.