C3 Submersible Fluorometer
998-2300 Rev. H Page 12
3.2.2 Current Data and Calibration Screen
The Current Data Screen allows users to view real-time data acquisition,
in both tabular or waveform display, as well as conduct calibrations for
each channel.
Current Data
Channel names are displayed but cannot be changed in the Current
Data screen.
Real-time values are displayed next to each channel and in the
adjacent Data Table.
A calibration button is available for each channel and the depth
sensor.
Real-time data can be saved by selecting the “Start Data Capture”
button. Users will be prompted to enter a file name and file location.
Data files will be saved as .csv format.
Tabular Screen
Real-time data including date, time, units and temperature are
continuously scrolling on the tabular screen for enabled channels.
Channel names will appear in the associated column headings.
If units are not selected during calibration, a relative fluorescence
units (RFU) label will appear.