C3 Submersible Fluorometer
998-2300 Rev. H Page 21
4. Maintenance and Warranty
4.1 Maintenance
4.1.1 Rinsing
The C3 Submersible Fluorometer should be rinsed or soaked in
freshwater following each deployment.
4.1.2 Care for the bulkhead connector
Install the 8-pin female end plug to the bulkhead between uses. A light
coat of Silicone spray should be used on the rubber of the male pins of
the bulkhead to aid in sealing. The manufacturer recommends 3M™
Silicone Lubricant Spray or Loctite 8021 spray. Note: You should
avoid using silicone grease. Do NOT use WD-40, it will destroy the
connectors.
4.2 Warranty Terms
Turner Designs warrants the C3 and accessories to be free from defects in
materials and workmanship under normal use and service for a period of 12
months from the date of shipment from Turner Designs with the following
restrictions:
• Turner Designs is not responsible for replacing parts damaged by accident
or neglect. Your instrument must be installed according to instructions in the
User’s Manual. Damage from corrosion is not covered. Damage caused by
customer modification of the instrument is not covered.
• This warranty covers only Turner Designs products and is not extended to
equipment used with our products. We are not responsible for accidental or
consequential damages, except in those states where this limitation is not
allowed. This warranty gives you specific legal rights and you may have
other rights which vary from state to state.
• Damage incurred in shipping is not covered.
Extended warranties are available for purchase.
Contact sales@TurnerDesigns.com