998-2300 Rev. H Page 37
Calibration Cup Use
With the C3 Submersible
Fluorometer’s Flow Cap Housing
installed:
1) Remove the End Cap.
2) Use a port plug to plug the port
located on the Flow Cap Housing.
3) Position the unit vertically with
optical head facing upward (see
diagram to the right).
In this position the Flow Cap Housing
can be used to hold standard
solutions for calibrating the C3
sensors.
Specifications
Max. Pressure 100 psi
Length 10.1 cm (3.98 in.)
Diameter 10 cm (3.94 in.)
Weight 455 g (1.003 lbs.)
Max. Volume 400 ml
Material Delrin
Port Plug