OPERATION
&
MAINTENANCE
MANUAL
Electronic Table -Top Autoclaves
models 1730, 2340, 2540, 3545, 3850, 3870E
2340, 3850, 3870 EA
2450EK
Cat. No. MAN205-0112-003E Rev. D
Br
inkmann Instruments, Inc., One Cantiague Road. P.O. Box 1019, Westbury, NY 11590-0207, (800) 645-3050, Fax: (516) 334-7506
Brinkmann Instruments (Canada) Ltd., 6670 Campobello Road, Mississauga, Ontario L5N 2L8, (800) 262-8715, Fax: (905) 826-5425

Questions and Answers

  • A
    Angela DunlapAug 31, 2025
    What causes 'High Pres' to be displayed on Tuttnauer 2450EK Laboratory Equipment?
    • K
      Kenneth GeorgeAug 31, 2025
      The 'High Pres' message on your Tuttnauer Laboratory Equipment can be due to a few reasons. It could mean the air jet is clogged and needs cleaning. Another possibility is that the heating elements are staying on instead of cycling. A faulty or dirty temperature sensor (PT100) could also be the cause. For the latter two issues, it's best to call a technician.
  • J
    jennifermillerSep 10, 2025
    What to do if Tuttnauer 2450EK Laboratory Equipment displays ‘Low Water’?
    • N
      Nicole AllenSep 10, 2025
      If the 'Low Water' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed with the gasket properly seated, and check that the machine is level. Replace the door gasket if there's any steam leakage.
  • A
    Ashley Ramos DDSAug 30, 2025
    How to clean a very dirty Tuttnauer Laboratory Equipment chamber?
    • C
      cody27Aug 30, 2025
      To clean a very dirty chamber in your Tuttnauer Laboratory Equipment, use Chamber Brite ™. Very dirty machines may require two cleanings.
  • D
    Danielle FoxSep 23, 2025
    Why does Tuttnauer Laboratory Equipment exhaust stage take so long?
    • E
      eileengarciaSep 23, 2025
      If the exhaust stage of your Tuttnauer Laboratory Equipment cycle is taking a very long time, it's normal if you are running a ‘liquids’ program. The slow exhaust can take between 15 and 20 minutes.
  • C
    Carrie RileySep 12, 2025
    What does ‘High Temp’ mean on my Tuttnauer Laboratory Equipment?
    • C
      Christian GarciaSep 12, 2025
      If the 'High Temp' message appears during heat up, it may indicate a faulty temperature sensor. If it occurs at other times, the heating elements might be staying on continuously. In either situation, it's recommended to call a technician for assistance with your Tuttnauer Laboratory Equipment.
  • S
    Sarah WoodwardSep 15, 2025
    What to do if Tuttnauer Laboratory Equipment chamber is not filling?
    • B
      Brianna SmithSep 15, 2025
      If your Tuttnauer Laboratory Equipment chamber is not filling, check the water level in the reservoir and for any obstructions blocking the pickup tube. Clean the water sensor and try filling the machine manually using the Manual Inlet key.
  • J
    Jeremy PhamSep 23, 2025
    What to do if Tuttnauer 2450EK Laboratory Equipment printer prints blank paper?
    • S
      Steven DavisSep 23, 2025
      If the printer of your Tuttnauer Laboratory Equipment prints, but nothing appears on the paper, ensure the paper is mounted correctly, as only one side is printable.
  • A
    Adam BranchAug 27, 2025
    What to do if Tuttnauer Laboratory Equipment process fails during heating stage with a HIGH TEMP message?
    • A
      Amy WilliamsAug 27, 2025
      If your Tuttnauer Laboratory Equipment process fails while initiating the heating stage and displays a HIGH TEMP message, try changing the setting of the thermostat.
  • J
    John HayesAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 2450EK?
    • D
      David JohnsonAug 26, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • A
    Austin Reynolds DDSAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer Laboratory Equipment?
    • M
      Melanie MitchellAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.