OPERATION
&
MAINTENANCE
MANUAL
Electronic Table -Top Autoclaves
models 1730, 2340, 2540, 3545, 3850, 3870E
2340, 3850, 3870 EA
2450EK
Cat. No. MAN205-0112-003E Rev. D
Br
inkmann Instruments, Inc., One Cantiague Road. P.O. Box 1019, Westbury, NY 11590-0207, (800) 645-3050, Fax: (516) 334-7506
Brinkmann Instruments (Canada) Ltd., 6670 Campobello Road, Mississauga, Ontario L5N 2L8, (800) 262-8715, Fax: (905) 826-5425

Questions and Answers

  • C
    christopherwrightSep 23, 2025
    What to do if Tuttnauer 2450EK Laboratory Equipment printer doesn't print?
    • J
      Jennifer Jensen MDSep 23, 2025
      If the printer on your Tuttnauer Laboratory Equipment isn't printing, first make sure the paper is inserted correctly. If the issue persists, switch off the machine and switch it back on while pressing the feed button on the printer. If a test printout appears, the printer is likely okay, and there may be an electronics issue. Contact your dealer to solve the problem. If no test printout occurs, there is a problem with the printer. Contact your dealer to solve the problem.
  • D
    Danielle FoxSep 23, 2025
    Why does Tuttnauer Laboratory Equipment exhaust stage take so long?
    • E
      eileengarciaSep 23, 2025
      If the exhaust stage of your Tuttnauer Laboratory Equipment cycle is taking a very long time, it's normal if you are running a ‘liquids’ program. The slow exhaust can take between 15 and 20 minutes.
  • A
    Angela DunlapAug 31, 2025
    What causes 'High Pres' to be displayed on Tuttnauer 2450EK Laboratory Equipment?
    • K
      Kenneth GeorgeAug 31, 2025
      The 'High Pres' message on your Tuttnauer Laboratory Equipment can be due to a few reasons. It could mean the air jet is clogged and needs cleaning. Another possibility is that the heating elements are staying on instead of cycling. A faulty or dirty temperature sensor (PT100) could also be the cause. For the latter two issues, it's best to call a technician.
  • A
    Anthony SmithAug 25, 2025
    What to do when my Tuttnauer 2450EK displays ‘High Temp’?
    • D
      Darius CoxAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the air jet and check for proper voltage at the wall outlet.
  • A
    Austin Reynolds DDSAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer Laboratory Equipment?
    • M
      Melanie MitchellAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • J
    John HayesAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 2450EK?
    • D
      David JohnsonAug 26, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • A
    Adam BranchAug 27, 2025
    What to do if Tuttnauer Laboratory Equipment process fails during heating stage with a HIGH TEMP message?
    • A
      Amy WilliamsAug 27, 2025
      If your Tuttnauer Laboratory Equipment process fails while initiating the heating stage and displays a HIGH TEMP message, try changing the setting of the thermostat.
  • D
    Darlene SullivanAug 27, 2025
    How to troubleshoot Tuttnauer 2450EK machine not responding?
    • J
      James BoyerAug 27, 2025
      If your Tuttnauer Laboratory Equipment machine isn't responding, verify that the main switch is in the ‘On’ position, the power cord is properly connected to both the machine and the mains, and the circuit breaker hasn't tripped. Also, check the reset button on the cut-out thermostat.
  • W
    William HaynesAug 27, 2025
    What should I do if my Tuttnauer Laboratory Equipment shows ‘Low Pres’?
    • S
      stanleystacyAug 28, 2025
      If the 'Low Pres' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed, the door gasket is properly seated, and there is no steam leakage. Check the leveling of the machine.
  • W
    wilsonamandaAug 28, 2025
    How to fix ‘High Pres’ error on Tuttnauer Laboratory Equipment?
    • G
      gwallaceAug 28, 2025
      If the 'High Pres' message appears on your Tuttnauer Laboratory Equipment, the air jet might be clogged; clean it. It could also mean that the heating elements are continuously on, or the temperature sensor is faulty. In these cases, contact a technician.

Tuttnauer 2450EK Specifications

General IconGeneral
BrandTuttnauer
Model2450EK
CategoryLaboratory Equipment
LanguageEnglish