Questions and Answers

  • J
    Joshua RiversAug 29, 2025
    How to unclog the drain on Tuttnauer Laboratory Equipment?
    • W
      wcombsAug 29, 2025
      If the drain of your Tuttnauer Laboratory Equipment is clogged: 1. Open the drain by turning it counterclockwise and place a heavy object over the reservoir cover. 2. Blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. 3. Alternatively, disassemble the drain valve, clean it, and blow out the line. Reassemble the valve, ensuring the large 'O' ring is properly seated.
  • A
    Ashley Ramos DDSAug 30, 2025
    How to clean a very dirty Tuttnauer Laboratory Equipment chamber?
    • C
      cody27Aug 30, 2025
      To clean a very dirty chamber in your Tuttnauer Laboratory Equipment, use Chamber Brite ™. Very dirty machines may require two cleanings.
  • H
    heathlydiaAug 30, 2025
    Why does Tuttnauer 2450EK Laboratory Equipment show 'High Temp'?
    • S
      Samantha DominguezAug 30, 2025
      If the 'High Temp' message appears on your Tuttnauer Laboratory Equipment during heat up, it could indicate a faulty temperature sensor. Alternatively, the heating elements might be staying on continuously instead of cycling on and off. In either situation, it's recommended to call a technician for assistance.
  • L
    lisa54Aug 25, 2025
    What should I do if my Tuttnauer 2450EK Laboratory Equipment displays ‘High Pres’?
    • M
      Mr. Michael Quinn IIIAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the air jet. Ensure the sterilization time and temperature are set correctly; if unsure, use the default settings. Also, make sure the autoclave is not too heavily loaded.
  • D
    danielle35Sep 1, 2025
    How to troubleshoot a clogged drain in Tuttnauer 2450EK Laboratory Equipment?
    • D
      Daniel PerezSep 1, 2025
      To address a clogged drain in your Tuttnauer Laboratory Equipment, open the drain by turning it counterclockwise. Place a heavy object over the reservoir cover, and blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. If needed, disassemble the drain valve by turning it counterclockwise past the stop point, remove the valve, clean it, and blow out the line. Reassemble the valve by turning clockwise, ensuring the large “O” ring is pressed into the groove on the autoclave.
  • G
    gdavisSep 1, 2025
    What to do if my Tuttnauer 2450EK Laboratory Equipment is not responding?
    • T
      Tony ArnoldSep 1, 2025
      If your Tuttnauer Laboratory Equipment is not responding, ensure the main switch is in the ‘On’ position. Also, verify the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and make sure the circuit breaker hasn't tripped; lift the circuit breaker lever if it has.
  • S
    Stacey RamosSep 2, 2025
    Why is my Tuttnauer 2450EK leaking at the door?
    • Z
      zcaseSep 2, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure the door is tightened enough and the door gasket is sealing the chamber. If necessary, replace the door gasket.
  • R
    Robert DuncanSep 3, 2025
    How to fix a Tuttnauer Laboratory Equipment chamber that is not filling?
    • S
      Sheri LoveSep 3, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is not filling, check the water level in the reservoir. Also, inspect the bottom of the reservoir for any obstruction blocking the pickup tube. Clean the water sensor and try filling the machine manually using the Manual Inlet key.
  • E
    Elizabeth WellsSep 3, 2025
    Why is the drying cycle taking so long in my Tuttnauer 2450EK Laboratory Equipment?
    • K
      karenkrauseSep 3, 2025
      If the drying cycle in your Tuttnauer Laboratory Equipment is taking too long, the autoclave may be overloaded. Also, consider replacing the HEPA filter.
  • A
    Angela JonesSep 4, 2025
    Why does the exhaust stage take so long when running a cycle on my Tuttnauer 2450EK Laboratory Equipment?
    • C
      Caleb NelsonSep 4, 2025
      If you are running a ‘glassware’ program on your Tuttnauer Laboratory Equipment, this is normal. The slow exhaust will take between 15 and 20 minutes.