TytoHome General Operation User Guide
Appendix D. Electromagnetic Compatibility (EMC) Information
Test
frequency
(MHz)
Bandᵃ⁾
(MHz)
Serviceᵃ⁾ Modulationᵇ⁾
Maximum
power (W)
)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Compliance
level
(V/m)
385 380 -
390
TETRA 400 Pulse mod-
ulationᵇ⁾ 18
Hz
1.8 0.3 27 27
450 430 -
470
GMRS 460,
FRS 460
FMᶜ⁾ ± 5 kHz
deviation 1
kHz sine
2 0.3 28 28
710 704 -
787
LTE Band
13, 17
Pulse mod-
ulationᵇ⁾ 217
Hz
0.2 0.3 9 9
745
780
810 800 -
960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse mod-
ulationᵇ⁾ 18
Hz
2 0.3 28 28
870
930
1720 1 700 -
1 990
GSM 1800;
CDMA
1900; GSM
1900;
DECT; LTE
Band 1, 3,
4, 25;
UMTS
Pulse mod-
ulationᵇ⁾ 217
Hz
2 0.3 28 28
1845
1970
2450 2 400 -
2 570
Bluetooth,
WLAN,
802.11
b/g/n, RFID
2450, LTE
Band 7
Pulse mod-
ulationᵇ⁾ 217
Hz
2 0.3 28 28
5240 5 100
- 5 800
WLAN
802.11 a/n
Pulse mod-
ulationᵇ⁾ 217
Hz
2 0.3 28 28
5500
5785
Table 3. Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless