TytoHome General Operation User Guide
Appendix D. Electromagnetic Compatibility (EMC) Information
Test
frequency
(MHz)
Bandᵃ⁾
(MHz)
Serviceᵃ⁾ Modulationᵇ⁾
Maximum
power (W)
)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Compliance
level
(V/m)
Note
: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the trans-
mitting antenna and the TytoCare Device may be reduced to 1 m. The 1 m test distance is per-
mitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while
it does not represent actual modulation, it would be worst case.
Table 3. Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment (continued)