Low power mode
The Minisonic II Portable allows energy safety through :
Power saving
OLED display turned o (switching on by pressing any keyor power button)
1 flow calculation per second (instead of 10 using standard mode)
This mode is the basic way to get power safety and keeping good flow calcultaion resolution
Menu
Measurement config.
Probes distance
US Signal display
Auto zero
Flowrate measurement
Measurement config.
Logger
Inputs / Outputs
Low power mode
Off
Mode
Power saving
Sequencer
Sequencer
Mesurement duration indicates how long is each periode of active and continuous flow calculation
Periode indicates how often the instrument wakes up
This mode allows the largest power safety option (depending on the settings)
Off
Mode
Power saving
Sequencer
Off
Mode
Power saving
Sequencer
Off
Mode
Power saving
Sequencer
Off
Mode
Power saving
Sequencer
Mode
Low power mode
Sequencer
Period
Low power mode
10 min
Measurement duration
Low power mode
10 s
Measurement duration
010 s
Min
: 5s
Max: 280s
Period
0015 min
Min:
5 min
Max : 1440 min
(....)
(....)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ultraflux Minisonic II Portable and is the answer not in the manual?

Ultraflux Minisonic II Portable Specifications

General IconGeneral
BrandUltraflux
ModelMinisonic II Portable
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals