Programming
75
Parameters to be set in a burst test:
• Test Pressure
• Couple Time
• Test Time
• Vent (0 or time)
• Max Burst (maximum burst level)
• Min Burst (minimum burst level)
• Burst Event (amount of pressure drop to be detected)
• Next Program
Minimum and Maximum for Burst
Minimum and maximum setpoints establish a window in which a
burst event must take place to be a pass. If the burst occurs below
the minimum setpoint, the test is stopped and LoBurst appears in
the status box on the run display. The fail light turns on.
If pressure goes above the maximum setpoint without the product
bursting, Sprint iQ stops the test and shows HiBurst in the status
box on the run display. The fail light turns on.
If product fails to reach a peak pressure above the maximum burst
setpoints before the end of the test phase, Sprint iQ stops the test
and shows NoBurst in the status box on the run display.