Operation errors ____________________________________________________________________
Vaisala _________________________________________________________________________ 32
CHAPTER 6
OPERATION ERRORS
The following indicates an operation error of the
HMT360:
•  Analog output(s) current drops below 4 mA
•  Display shows ERROR 
•  LED on the cover is lit and no calibration is going on (transmitters
without a display)
In the case of error:
•  Check that the sensor is connected properly
•  Check if there is condensed water in the probe, if yes, let the probe dry 
Cleaning the sensor from glycol in a natural
gas application
If a glycol layer on the VAISALA HUMICAP
®
 sensor is disturbing
the moisture measurement, the sensor can be cleaned on site soaking
the sensor to de-ionized water or to IPA (isopropanol, propan-2-ol).