Operation errors ____________________________________________________________________
Vaisala _________________________________________________________________________ 32
CHAPTER 6
OPERATION ERRORS
The following indicates an operation error of the
HMT360:
Analog output(s) current drops below 4 mA
Display shows ERROR
LED on the cover is lit and no calibration is going on (transmitters
without a display)
In the case of error:
Check that the sensor is connected properly
Check if there is condensed water in the probe, if yes, let the probe dry
Cleaning the sensor from glycol in a natural
gas application
If a glycol layer on the VAISALA HUMICAP
®
sensor is disturbing
the moisture measurement, the sensor can be cleaned on site soaking
the sensor to de-ionized water or to IPA (isopropanol, propan-2-ol).