Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 43
Stopping the continuous output
S ↵
The continuous outputting is stopped with the command S. If
outputting is active, this command is not echoed.
Setting the output interval
INTV n xxx ↵
where: n = 1 - 255 xxx = S, MIN or H
Sets the output interval when the transmitter outputs measurement
readings. The time interval is used when the continuous output is
active. For example, the output interval is set to 10 minutes:
>INTV 10 min ↵
Output intrv. : 10 min
>
Setting the ambient pressure for calculations
Turn the internal dip switch CALIBRATION ENABLED/DISABLED to
ON-position before setting the pressure and return it to OFF- position
after setting.
PRES aaaa.a ↵
where: aaaa.a = pressure (hPa)
>PRES 2400 ↵
Pressure : 2400
>
Pressure conversion chart
mmHg/Torr inHg atm bar psi
hPa/mbar
1 1.333224 33.86388 1013.25 1000 68.94757
mmHg/Torr
0.7500617 1 25.40000 760 750.0617 51.71493
inHg
0.02952999 0.03937008 1 29.921 29.52999 2.036021
atm
0.00098692 0.00131597 0.033422 1 0.98692 0.068046
bar
0.001 0.001333224 0.03386388 1.01325 1 0.06894757
psi
0.01450377 0.01933678 0.4911541 14.6962 14.50377 1
Example: 29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa / mbar
NOTE
Conversions from mmHg and inHg are defined at 0 °C and for
mmH
2
O and inH
2
O at 4 °C.
FROM
TO