Serial interface _____________________________________________________________________
Vaisala _________________________________________________________________________ 42
Output commands
Testing the analog outputs
This command outputs the current value of each channel and the
corresponding control signal of the digital-to-analog converter.
ITEST aa.aaa bb.bbb
where: aa.aaa = current value to be set for channel 1 (mA)
bb.bbb = current value to be set for channel 2 (mA)
For example:
>ITEST 8 12
8.000 3F8 12.000 70O
>
>ITEST
6.349 25A 19.001 E93
>
The set current values remain valid until you give the command
ITEST without readings. With this command the desired outputs of
the transmitter are shown.
Note! When outputting low currents from channel 1, remember to
remove RS232C-cable while reading the current output, because of
the increased current consumption for using RS port.
Outputting the measurement
values
SEND
This command outputs the measured values in one point.
Activating a continuous output
R
With the command R the transmitter outputs measured values
continuously.