Display/keypad commands____________________________________________________________
Vaisala _________________________________________________________________________ 36
Press button C on the display cover: text "SCAL" appears on the
display. Then press button E until the following display appears:
The characters on the second line of the display in this menu
correspond to the quantities according to the following table:
Table 2 Available Quantities
character  quantity abbreviation availability
0 relative humidity RH A, D
1 temperature T A, D, F, H, N
2 dewpoint temperature Td D, N
3 absolute humidity a D
4 mixing ratio x D
5 wet bulb temperature Tw D
6
volume concentration
ppm H, N
7 water activity aw H, F
8 relative humidity of saturation RS H
9 saturation temperatur Ts H
A water content w N
A: available for HMT360A
D: available for HMT360D
F: available for HMT360F
H: available for HMT360H
N: available for HMT360N
Select the quantity for channel 1 with buttons ▲ and ▼ and
acknowledge the selection with button E.
If the transmitter is equpped with two analog channels, select the
quantity for channel 2 in the same way, for example:
Press button C to exit the display command mode or continue by
setting the pressure.
NOTE
Remember to restore the dip switch settings.