REM Alarm (two tones)
5-2 Force 2 User’s Guide
2. Inspect the patient REM pad. Do not remove the pad.
a. Apply firm pressure over the entire surface area of the patient
return electrode, particularly the center.
b. If the REM alarm persists, place the generator in the standby mode.
c. Unplug the patient return electrode from the generator. Do not
remove the return electrode pad from the patient.
d. Proceed to step 3.
3. Apply a second patient REM pad.
a. Apply a second REM electrode to the patient's thigh, upper biceps,
calf, or lower back.
b. Place the generator in the ready mode.
c. Connect the return electrode to the generator.
d. If the alarm condition is not corrected, place the generator in the
standby mode.
e. Unplug the patient return electrode from the generator. Do not
remove the return electrode from the patient.
f. Proceed to step 4 or 5.
Original application
Application 2