10. Capture the CT Number phantom and evaluate CT Number accuracy.
(refer to 5. CT Number Accuracy evaluation.)
※ CT Number standard value
Parameter LSL(Lower Specification Limit) USL(Upper Specification Limit)
Air -1030 -900
Water -20 20
Teflo n(Bone) 900 1100
11. If CT Number does not meet the standard as above, adjust the CT Number value.