EasyManua.ls Logo

Vatech A9

Vatech A9
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
151 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
2.2. PA Mode Start Position Setting
1. Attach Cu 0.8T in front of the collimator.
2. Attach the CEPH standard phantom to the ear rod
3. Click the Console software as shown in the picture below to capture the Normal PA Mode.
1. Choose the CEPH.
2. Choose the High
3. Choose the PA.
4. Choose the Confirm button.
5. Choose the Ready button.

Table of Contents

Other manuals for Vatech A9

Related product manuals