5.2. Test Method
1. Complete the collimator set up and acquire the image in the stopped state.
2. Run C:\VCaptureSW\Utility\EzEval_Phantom\EzEval.exe program.
3. Enter 90% as the threshold standard and check if it satisfies the Collimator Alignment standard.
5.3. Test Result
High-Resolution Lateral Mode
Parameter LSL Measurement value USL Pass/Fail
Threshold:
90%
Upper 20 40
Lower 20 40
Left 1 5
Right 1 5